Biomedical Medical Center (Northern Taiwan)

BSC Performance Verification & BSL-3 Laboratory Audit

In support of a novel respiratory pathogen research program, a northern medical center required rapid performance verification of 30 Class II biosafety cabinets (BSCs) and the overall BSL-3 facility, together with audit recommendations. SYE deployed NSF/ANSI 49 accredited technicians and completed all verification and audit deliverables within four weeks.

醫學中心生物安全驗證
Challenge

Challenge

A full performance report was required before the BSL-3 laboratory could go live. The 30 BSCs spanned five manufacturers with varying service ages; for several legacy units the OEM technical documentation was lost, so baseline parameters had to be re-established on site.

Approach

Our Approach

  1. 01

    Unit-by-unit NSF/ANSI 49 Enhanced Accreditation testing: airflow, HEPA, smoke pattern, UV intensity

  2. 02

    BSL-3 envelope pressure gradient (≥ -50 Pa), air change rate, and airflow visualization

  3. 03

    On-site audit covering personnel flow, waste handling, PPE storage and usage SOPs

  4. 04

    Non-conformance reports with remediation recommendations and re-test scheduling support

Outcomes

Outcomes

  • All units pass NSF/ANSI 49 verification
  • Pressure gradient meets BSL-3 spec
  • Full audit report delivered